6% Ruhr Joint Negative Pressure Leakage Rate Test Machine
GET A QUOTESpecification of 6% Ruhr Joint Negative Pressure Leakage Rate Test Machine
Item | Description |
Executive standard | Comply with the relevant standards of ISO80369-20, as well as the standards of GB15810Disposable Sterile Syringes and ISO7886-1 |
Test items | Negative pressure leakage rate test of Ruhr joint, negative pressure for syringe sealing, negative pressure for infusion device sealing |
Nominal capacity | Any choice. |
Negative pressure detection | -10~-94Kpa, error not greater than ± 1% |
Resolution | 0.01 kpa |
Negative pressure leakage rate resolution | -0.0001Pa.m3/s |
Test time | 1s~999s,arbitrarily set(the negative pressure of the measurement system should not drift ± 0.5Kpa within 5 minutes) |
Power off storage | Yes |
Overall power | 200 W |
Sensor | High-precision force sensor (with software and hardware protection function) |
Fixture | Stainless steel (easy to grip) |
Power supply | AC 220 V / 50 Hz |
Force measuring mechanism | Built-in |
Printer | Airborne needle printer |
Printing method | Test results can be printed |
Data storage | USB flash drive can be imported into the computer for storage |
Do you like our products?
You can click the "Get a Quote" button and fill in the form, We will contact you promptly.